Chapter 2 Command Summary
HP 53131A/132A Command Summary
2-28 Programming Guide
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax Parameter Form Std/
New
Comments
:DIAGnostic (Cont.)
:CALibration (Cont.)
:ROSCillator
:AUTO
:STATus?
:TINTerval
:FINE
:QUICk
ONCE | OFF
[1 | 2 | 3 | 4]
New
New
New
New
New
New
Subtree. ROSCillator is an abbreviation for
Reference OSCillator.
ONCE calibrates the timebase. This command is
usable only if the instrument contains the medium or
high stability oscillator option.
Query only. Returns status of last calibration.
0 = pass; 1 = fail.
Subtree.
Event; no query. Four steps which calibrate out
Ch1, 2 electrical path length differences.
Event; no query. Calibrates out Ch1, 2 electrical path
length differences.
:DISPlay
:ENABle
:MENU
[:STATe]
[:WINDow]
:TEXT
:FEED
:RADix
<Boolean>
OFF
“[:]CALCulate2” |
“[:]CALCulate3”
COMMa | DPOint
Std
Std
Std
Std
Std
Std
Std
New
Subsystem. Controls the selection and presentation
of textual information on the display.
Controls whether the whole display is visible.
Subtree.
Sets the Counter to switch from the menu display to
the result display.
Subtree.
Subtree. Allows for the display of textual information.
Sets which data flow is fed into the display. “CALC2”
specifies the raw measurement, scaled/offset
measurement, or Limit Graph display. “CALC3”
specifies the statistics result display.
Sets the character used to separate integral and
fractional portions of a number. (USA numerical
convention is Decimal POint.)
:FETCh
Std See Measurement Instructions in this table.
:FORMat
[:DATA]
ASCii | REAL
Std
Std
Subsystem. Sets a data format for transferring
numeric information.
Sets the data format.
:HCOPy
:CONTinuous
<Boolean>
New
New Enables or disables printing results.