Chapter 4 Command Reference
Introduction
Programming Guide 4-33
:DIAGnostic:CALibration:INPut[1|2]:OFFSet:
AUTO ONCE | OFF
Calibrates the channel 1 or 2 input trigger OFFSet when the ONCE parameter is
used.
Before sending this command, BE SURE to disconnect any input signal from the
appropriate input.
A sequence of ASCII-encoded bytes: OFF
• The calibration values are stored in non-volatile memory, and are unaffected
by power-on, save/recall, and *RST.
• Use :DIAG:CAL:STAT? to check for successful calibration.
• After calibration is completed, the state of this command’s parameter is OFF.
Scale and Offset/POWER (Calibration menu).
:DIAGnostic:CALibration:INTerpolator:AUTO ONCE | OFF |ON
Calibrates the interpolator circuit in the Counter when the ONCE parameter is
used.
AUTO ON enables automatic interpolator calibration on every measurement.
AUTO OFF disables automatic interpolator calibration.
A sequence of ASCII-encoded bytes: OFF or ON
• *RST: ON
• Use :DIAG:CAL:STAT? to check for successful calibration.
• After ONCE calibration is completed, the state of this command’s parameter
is OFF.
• When :DIAG:CAL:INT is set to OFF, the Counter reports the positive status
condition (questionable Time, Frequency, and Phase) to bits 2, 5, and 6 of the
Questionable Status Register. When :DIAG:CAL:INT is set to ON, the
Counter reports the negative status condition (NOT questionable Time,
Frequency, and Phase) to bits 2, 5, and 6 of the Questionable Status Register.
• This enable is unaffected by save/recall.
Query Response
Comments
Related
Front-Panel
Keys
Query Response
Comments