Chapter 2 Command Summary
HP 53131A/132A Command Summary
2-26 Programming Guide
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax Parameter Form Std/
New
Comments
:CALCulate2 (Cont.)
:LIMit (Cont.)
:FCOunt
:LOWer?
:UPPer?
[:TOTal]?
:LOWer
[:DATA]
:STATe
:UPPer
[:DATA]
:PCOunt
[:TOTal]?
<numeric_value> [HZ | S
| DEG]
<Boolean>
<numeric_value> [HZ | S
| DEG]
Std
New
New
New
Std
Std
Std
Std
Std
New
New
Subtree. An abbreviation for Fail COunt.
Query only. Returns the number of limit test
failures at the lower limit.
Query only. Returns the number of limit test
failures at the upper limit.
Query only. Returns the total number of
measurements that failed the limit test.
Subtree.
Sets lower limit used in limit testing.
Sets the limit test enable. Note that this setting
must be enabled for any of the other :CALC2
settings can be used.
Subtree.
Sets upper limit used in limit testing.
Subtree. An abbreviation for Pass COunt.
Query only. Returns the total number of
measurements that passed the limit test.
:CALCulate3
:AVERage
:ALL?
:CLEar
:COUNt
:CURRent?
[:STATe]
:TYPE
:DATA?
:FEED
<numeric_value>
<Boolean>
MAXimum | MINimum |
SDEViation | SCALar or
MEAN
“[:]CALCulate[1] ”
Std
Std
New
Std
Std
New
Std
Std
Std
Std
Subsystem. Performs post-acquisition statistics
computation and data transfer.
Subtree. Collects together the commands
associated with the Statistics capabilities.
Returns all four Statistics results (i.e., mean,
standard deviation, maximum, and minimum).
Event; no query. Clears the statistics results
and statistics count.
Selects number of measurements to combine
for statistics.
Query only. Returns the current number of data
values collected, thus far.
Enables/disables statistics post-processing.
Note that this setting must be enabled for any
of the other :CALC3 settings to be used.
Selects which statistic will be in
:CALC3:DATA?,and on the front-panel display.
Query only. Returns statistic result specified by
:CALC3:AVER:TYPE.
Sets the data flow to be fed into the
CALCulate3 block.