HP (Hewlett-Packard) 132A Water Dispenser User Manual


 
Chapter 4 Command Reference
Introduction
Programming Guide 4-85
[:SENSe]:EVENt3:SLOPe?
Queries which edge of channel 3 input port will be considered an event.
A sequence of ASCII-encoded bytes: POS
[:SENSe]:FREQuency Subtree
This subtree controls the Frequency, Frequency Ratio, and Period measuring
capabilities of the instrument.
[:SENSe]:FREQuency:ARM Subtree
This subtree is used to synchronize the Frequency, Frequency Ratio, and Period
start and stop arm with events. The following combination of start/stop arming
sources are valid:
STARt:SOURce STOP:SOURce Front-Panel Gating Settings
GATE STOP
IMMediate IMMediate AUTO _______
IMMediate TIMer TIME _______
IMMediate DIGits DIGITS _______
EXTernal EXTernal EXTERNAL POS, NEG
EXTernal TIMer EXTERNAL TIME
EXTernal IMMediate EXTERNAL AUTO
[:SENSe]:FREQuency:ARM[:STARt]:SLOPe POSitive | NEGative
Sets or queries the slope of the external start arm signal used in external arming
Frequency, Frequency Ratio, and Period measurements.
A sequence of ASCII-encoded bytes: POS or NEG
*RST: POSitive
Only applies when [:SENS]:FREQ:ARM[:STAR]:SOUR EXT is selected.
Gate & ExtArm
Query Response
Query Response
Comments
Related
Front-Panel
Keys