Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Programming Guide 2-19
Calibration Menu to SCPI Command Map (Cont.)
1 a. :CALibration:SECurity:STATe?
b. :DIAGnostic:CALibration:INPut1:OFFSet:AUTO ONCE
:DIAGnostic:CALibration:INPut2:OFFSet:AUTO ONCE
:DIAGnostic:CALibration:INPut1:GAIN:AUTO ONCE
:DIAGnostic:CALibration:INPut2:GAIN:AUTO ONCE
:DIAGnostic:CALibration:TINTerval:QUICk
:DIAGnostic:CALibration:TINTerval:FINE[1 | 2 | 3 | 4]
:DIAGnostic:CALibration:ROSCillator:AUTO ONCE
c
1
. :CALibration:SECurity:CODE <new_code>
OR
:CALibration:SECurity:STATe ON, <present_code>
c
2
. :CALibration:SECurity:STATe OFF, <present_code>
d. :CALibration:COUNt?
e. No command
The Calibration Menu is accessed by holding the Scale & Offset key and cycling
POWER key.
Figure 2-6. Calibration Menu to SCPI Command Map (Part 2 of 2)