Chapter 4 Command Reference
[:SENSe] Subsystem
4-100 Programming Guide
[:SENSe]:TINTerval:ARM:ESTART:LAYer2:SLOPe POSitive |
NEGative
Sets or queries the slope of the external start arm signal used in external arming
time interval (including Time Interval, Risetime, Falltime, Duty Cycle, and Pulse
Width functions) measurements.
A sequence of ASCII-encoded bytes: POS or NEG
• *RST: POSitive
• Only applies when [:SENS]:TINT:ARM:ESTART:LAY2:SOUR EXT is
selected.
Gate & ExtArm
[:SENSe]:TINTerval:ARM:ESTART:LAYer2:SOURce IMMediate |
EXTernal
Sets or queries the start arm for time interval (including Time Interval, Risetime,
Falltime, Duty Cycle, and Pulse Width functions) measurements.
A sequence of ASCII-encoded bytes: IMM or EXT
*RST: IMMediate
Gate & ExtArm
[:SENSe]:TINTerval:ARM:ESTART[:LAYer[1]]:ECOunt
<numeric_value>
Sets or queries the number of events used to delay the start arm for Time Interval
measurements.
1 to 99,999,999
1
Numeric data transferred as ASCII bytes in <NR1> format.
• *RST: 1
• Only applies when [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:SOUR
INTernal1 is selected.
Query Response
Comments
Related
Front-Panel
Keys
Query Response
Comments
Related
Front-Panel
Keys
<numeric_value>
Range
<numeric_value>
Resolution
Query Response
Comments