Chapter 4 Command Reference
[:SENSe] Subsystem
4-96 Programming Guide
[:SENSe]:TINTerval:ARM[:STARt]:SOURce IMMediate | EXTernal
Sets or queries the start arm for time interval (including Time Interval, Risetime ,
Falltime, Duty Cycle, and Pulse Width functions) measurements.
A sequence of ASCII-encoded bytes: IMM or EXT
*RST: IMMediate
Gate & ExtArm
[:SENSe]:TINTerval:ARM:STOP:SOURce IMMediate | TIMer
Sets or queries the stop arm (that is, the delay) for Time Interval measurements.
A sequence of ASCII-encoded bytes: IMM or TIM
• *RST: IMMediate
• This [:SENS]:TINT:ARM:STOP:SOUR command has no affect on the
following measurements: Risetime, Falltime, Duty Cycle, and Pulse Width.
Gate & ExtArm
[:SENSe]:TINTerval:ARM:STOP:TIMer
<numeric_value> [S]
Sets or queries the time used to delay the stop arm for Time Interval
measurements.
• For short gate time: 0.1E-3* to 99.99E-3 seconds
• For long gate time: 10 0E-3 to 10.000 seconds
• For short gate time: 0.01E-3 seconds
• For long gate time: 1E-3 seconds
*For HP 53131A with firmware revisions below 3427, the minimum gate time is 1 ms.
*For HP 53132A with firmware revisions below 3646, the minimum gate time is 1 ms.
Query Response
Comments
Related
Front-Panel
Keys
Query Response
Comments
Related
Front-Panel
Keys
<numeric_value>
Range
<numeric-value>
Resolution