HP (Hewlett-Packard) 132A Water Dispenser User Manual


 
Chapter 4 Command Reference
Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)
4-78 Programming Guide
The work-around commands which should be substituted are listed below.
:INIT; *WAI
:FETCh:MAXimum?
:INIT; *WAI
:FETCh:MINimum?
:INIT; *WAI
:FETCh:PTPeak?
The command listed below should NOT be used when the function choice is
Voltage Peaks.
:READ?
The work-around command sequence which should be substituted is listed below.
:INIT; *WAI
:FETCh?
When the function is set to Voltage Peaks, and the Counter is in SINGLE mode,
the HP-IB command :FETCh? will hang the HP-IB if the measurement has not
completed.
The work around inserts a command which enforces correct timing. The
work-around command sequence is listed below.
:INIT; *WAI
:FETCh?
When the function is set to Voltage Peaks, and the Counter is in RUN mode, the
HP-IB command :FETCh? should not be used.
The work around puts the box into SINGLE mode, and inserts a command which
enforces correct timing. The work-around command sequence is listed below.
:INIT:CONTinuous OFF
:INIT; *WAI
:FETCh?