Agilent Technologies Agilent 86120C Ventilation Hood User Manual


 
4-17
Programming Commands
Measurement Instructions
MEASure{:ARRay | [:SCALar]} :POWer?
Returns amplitude values.
Syntax
:POWer? [<expected_value>[,<resolution>]]
Description When used with a :SCALar command, a single value is returned. The
display is placed in the single-wavelength mode, and the marker is
placed on the signal having a power level that is closest to the
<expected_value> parameter.
When used with an :ARRay command, an array of amplitudes is
returned. The display is placed in the list-by-power mode.
Returned values are in the current power units. Wavelength units are
not affected.
<expected_value>
Constants
MAXimum Displays the highest power signal
MINimum Displays the lowest power signal
DEFault Displays the signal at the current marker position
Used With <expected_value> <resolution>
SCALar optional ignored
ARRay ignored ignored
CONFigure command
When this function is used with the CONFigure command, the query question
mark character ? must not be included in the string. However, the FETCh, READ,
and MEASure command are queries and require the question mark. Refer to the
examples for this command.