4-78
Programming Commands
DISPlay Subsystem
MARKer:MAXimum:RIGHt
Moves the marker right to the next laser line.
Syntax
:DISPlay:MARKer:MAXimum:RIGHt
Attribute
Summary
Preset State: marker set to maximum-power laser line
*RST State: marker set to maximum-power laser line
SCPI Compliance: instrument specific
Command Only
Description Moves the marker from the current marker position to the next laser
line having the following characteristic:
•longer wavelength
•higher frequency
•higher wave number
If the display is in the List by Ampl mode, it will be changed to List by
WL before the marker is moved.
[WINDow]:GRAPhics:STATe
Turns on and off the display of the power bars.
Syntax
:DISPlay[:WINDow]:GRAPhics:STATe{?| {ON | OFF | 1 | 0}}
Attribute
Summary
Preset State: on
*RST State: on
SCPI Compliance: standard
Description Specifying on displays the power bars in all modes except the drift
and signal-to-noise modes. Specifying off prevents the display of
power bars for all instrument modes.