4-62
Programming Commands
CALCulate3 Subsystem
FPERot[:STATE]
Turns on and off the Fabry-Perot measurement mode.
Syntax
:CALCulate3:FPERot[:STATE] {? | {ON | OFF | 1 | 0}}
Attribute
Summary
Preset State: off
*RST State: off
SCPI Compliance: instrument specific
Description When the state is ON, the Agilent 86120C measures characteristics of
the Fabry-Perot laser modes. The modes are defined by the peak
excursion and peak threshold commands.
FPERot:FWHM?
Queries the full width half-maximum data of the selected modes.
Syntax
:CALCulate3:FPERot:FWHM{[:WAVelength] | :FREQuency | :WNUMber}?
Argument Description
WAVelength Returns the full width half-maximum wavelength of the selected
modes.
FREQuency Returns the full width half-maximum frequency of the selected
modes.
WNUMber Returns the full width half-maximum wavenumber of the selected
modes.
Example Query
Response
WAVelength
+5.47128800E009
FREQuency +6.93436400E+011
WNUMber +2.31306200E+001