Agilent Technologies Agilent 86120C Ventilation Hood User Manual


 
4-99
Programming Commands
SYSTem Subsystem
HELP:HEADers?
Queries a listing of all the remote programming commands available
for the Agilent 86120C.
Syntax
:SYSTem:HELP:HEADers?
Attribute
Summary
Preset State: none
*RST State: none
SCPI Compliance: instrument specific
Query Only
Description The returned ASCII string of commands is in the IEEE 488.2 arbitrary-
block data format. The first line indicates the total number of bytes
returned to the computer. That is, the # character is followed by one
digit which indicates how many of the following digits convey the byte
count. The next digits give the actual byte count. For example, in the
listing below, 4387 bytes are indicated in the file.
Each command in the listing is separated by a linefeed character.
The following is an example of the first few lines and last few lines
returned in the string. The term nquery indicates that a command
cannot be sent as a query. The term qonly indicates that a command
can only be sent as a query.
#44387
:ABORt/nquery/
:CALCulate:DATA?/qonly/
:CALCulate:TRANsform:FREQuency:POINts
:CALCulate1:DATA?/qonly/
:CALCulate1:TRANsform:FREQuency:POINts
:CALCulate2:DATA?/qonly/
.
.
.
*IDN?/qonly/
*OPC
*RCL/nquery/
*RST/nquery/
*SAV/nquery/
*SRE
*STB?/qonly/
*TRG/nquery/
*TST?/qonly/
*WAI/nquery/