Agilent Technologies Agilent 86120C Ventilation Hood User Manual


 
4-43
Programming Commands
CALCulate2 Subsystem
WLIMit:STOP:WNUMber
Sets the stopping wavenumber for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP:WNUMber {?|{<real> | MINimum | MAXimum }}
<real> is a wavenumber value that is within the following limits:
Constant
Description
MINimum start wavelength limit
MAXimum 7874 cm
-1
(1270 nm)
Attribute
Summary
Non-sequential command
Preset State: 7.87401E5 m
-1
*RST State: 7.87401E5 m
-1
SCPI Compliance: instrument specific
Description This command sets the stopping range for the wavelength limit. The
default units for the <real> parameter are inverse meters. The stop
wavenumber value must be less than or equal to the start wavenumber
value or the stop wavenumber will be clipped to the start wavenumber
and a “Data out of range” error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument. Refer
to Always force the Agilent 86120C to wait for non-sequential commands on
page 3-12 for more information.