Agilent Technologies Agilent 86120C Ventilation Hood User Manual


 
4-42
Programming Commands
CALCulate2 Subsystem
WLIMit:STOP[:WAVelength]
Sets the stopping wavelength for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP[:WAVelength] {?|{<real> | MINimum | MAXimum }}
<real> is a wavelength value that is within the following limits:
Constant
Description
MINimum start wavelength limit
MAXimum 1650.0 nm
Attribute
Summary
Non-sequential command
Preset State: 1650 nm
*RST State: 1650 nm
SCPI Compliance: instrument specific
Description This command sets the stopping range for the wavelength limit. The
default units for the <real> parameter are meters. The stop wavelength
value must be greater than or equal to the start wavelength value or
the stop wavelength will be clipped to the start wavelength and a
“Data out of range” error will be generated. Setting the start wave-
length is equivalent to setting the start frequency/wavenumber because
of the inverse relationship of frequency to wavelength.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument. Refer
to Always force the Agilent 86120C to wait for non-sequential commands on
page 3-12 for more information.