Agilent Technologies Agilent 86120C Ventilation Hood User Manual


 
4-21
Programming Commands
Measurement Instructions
MEASure{:ARRay | [:SCALar]} :POWer:WAVe-
length?
Returns wavelength values.
Syntax
:POWer:WAVelength? [<expected_value>[,<resolution>]]
Description When used with a :SCALar command, a single value is returned. The
display is placed in the single-wavelength mode, and the marker is
placed on the signal having a wavelength that is closest to the
<expected_value> parameter. Default units for <expected_value>
parameter are in meters.
When used with an :ARRay command, an array of wavelengths is
returned. The display is placed in the list-by-wavelength mode.
The <resolution> parameter sets the resolution of the measurement. It
is a unitless number whose value will be limited to either 0.01 or
0.001 (whichever is closer).
Returned values are in meters. Displayed units are nanometers. Power
units are not affected.
<expected_value>
Constants
MAXimum The highest wavelength signal
MINimum The lowest wavelength signal
DEFault The current marker position
Used With <expected_value> <resolution>
SCALar optional optional
ARRay
ignored
a
a. Although ignored, this argument must be present if the resolution argument is specified.
optional
CONFigure command
When this function is used with the CONFigure command, the query question
mark character ? must not be included in the string. However, the FETCh, READ,
and MEASure command are queries and require the question mark. Refer to the
examples for this command.