Agilent Technologies E1441A Oven User Manual


 
102 Agilent E1441A SCPI Command Reference
Chapter 3
BURST MODULATION COMMANDS
Use the APPLy command or the equivalent FUNC:SHAP, FREQ, VOLT, and
VOLT:OFFS commands to configure the carrier waveform. Set the carrier frequency
between 100 µHz and 5 MHz (100 kHz maximum for triangle and ramp). The default
is 1 kHz.
BM:NCYCles
[SOURce:]BM:NCYCles <# cycles>|INFinity|MIN|MAX sets the number of cycles
to be output per burst (counted burst mode only). Select from 1 cycle
to 50,000 cycles, in 1 cycle increments. The default is 1 cycle.
MIN = 1 cycle. MAX =
50,000 cycles. [ Stored in volatile memory ]
Comments For sine, square, and arbitrary waveforms (does not apply to ramp and triangle
waveforms), the relationship between the carrier frequency and the minimum
burst count is shown below.
If you attempt to set the carrier frequency to a value that is not valid,
the function generator will automatically adjust the frequency to the maximum
value allowed with the present burst count. A -221, “Settings conflict” error is
generated and the carrier frequency is adjusted.
For all waveforms used with burst, if the carrier frequency is set less than or
equal to 100 Hz, the following relationship applies.
If you attempt to set the carrier frequency to a value that is not valid, the
function generator will automatically adjust the frequency to the minimum
value allowed with the present burst count. A -221, “Settings conflict” error is
generated and the frequency is automatically adjusted.
BM:NCYCles?
[SOURce:]BM:NCYCles? [MIN|MAX] queries the burst count. Returns an integer
between 1 and 50,000, or +9.90000E+37 (in the case of INF).
Carrier Frequency Minimum Burst Count
10 mHz to 1 MHz 1
>1 MHz to 2 MHz 2
>2 MHz to 3 MHz 3
>3 MHz to 4 MHz 4
>4 MHz to 5 MHz 5
BurstCount
CarrierFrequency
-------------------------------------------------
500 second