Agilent Technologies E1441A Oven User Manual


 
148 Agilent E1441A Error Messages
Appendix B
702 Cal secured
The function generator is secured against calibration.
703 Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the function generator. You must use the same security
code to unsecure the function generator as was used to secure it, and vice
versa. The security code may contain up to 12 alphanumeric characters. The
first character must be a letter.
704 Secure code too long
A security code was received which contained more than 12 characters.
705 Cal aborted
A calibration in progress is aborted when you press any front-panel key,
send a device clear, or change the local/remote state of the instrument.
706 Cal value out of range
The specified calibration value (CAL:VALUE) is invalid for the presently
selected function.
707 Cal signal measurement out of range
The specified calibration value (CAL:VALUE) does not match the signal
applied to the function generator.
708 Flatness cal failed
760 RAM checksum failure
770 Nonvolatile arb waveform memory checksum failure
771 Nonvolatile cal memory checksum failure
772 Nonvolatile system memory checksum failure
This error may occur at power-up if memory location "0" has been deleted
(MEM:STAT:DEL 0 executed). Use *SAV 0 to store any valid state.
Please see the “MEMory:STATe:RECall:AUTO OFF | ON | 0 | 1” command
on page 85 for details on using the *SAV 0 state for an alternate power-up
state
773 Nonvolatile state memory checksum failure
774 Nonvolatile memory erase failure
775 Nonvolatile memory write failure
780 to 788 See “Arbitrary Waveform Errors” on page 149.
800 Block length must be even
The function generator represents binary data for arbitrary waveforms as a
12-bit integers, which is sent as two bytes (DATA:DAC VOLATILE