Agilent Technologies E1441A Oven User Manual


 
Agilent E1441A Error Messages 147
Appendix B
command to enable this error.
800 Block length must be even
The function generator represents binary data for arbitrary waveforms as a
12-bit integers, which is sent as two bytes (DATA:DAC VOLATILE
command only). An odd number of bytes is not accepted because the
function generator would not know how to interpret the last single byte.
810 State has not been stored
The memory location specified in the *RCL command was not used in a
previous *SAV command. You cannot recall the instrument state from a
memory location that was not previously specified as a storage location.
Self-Test Errors
The following errors indicate failures that may occur during a self-test. Refer
to the Service Guide for more information.
590 I/O processor reset
601 Trigger test failed
602 RAM read/write failed
603 Waveform RAM readback failed
604 Modulation RAM readback failed
606 Waveform ASIC failed
607 SYNC signal detection failure, Bessel filter path
608 SYNC signal detection failure, Elliptic filter path
625 I/O processor does not respond
626 I/O processor failed self-test
Calibration Errors
The following errors indicate failures that may occur during a calibration.
Refer to the Service Guide for more information.
701 Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the
function generator. When applicable, this error will occur at power-on to
warn you that the function generator is unsecured.