193
Appendix D
Table D-26. Performance Test Record for the Agilent E1441A (Page 3 of 7)
Agilent E1441A Frequency Accuracy
Function Output Minimum Measured Maximum M.U. TAR
Sine wave 3.5V @ 1.0 kHz 999.99 Hz 1000.01 Hz 0.55e-3 >10:1
Square wave 3.5V @ 1.0 kHz
500 Hz burst rate
495 Hz 505 Hz 1e-4 >10:1
Agilent E1441A Function Gain and Linearity
Function Output Minimum Measured Maximum M.U. TAR
Sine wave 7.0V @ 1.0 kHz 6.93 Vrms 7.07 Vrms 6.5e-3 >10:1
Sine wave 5.7V @ 1.0 kHz 5.643 Vrms 5.757 Vrms 5.9e-3 9.7:1
Triangle wave 5.7V @ 100 Hz 5.643 Vrms 5.757 Vrms 5.9e-3 9.7:1
Ramp wave 5.7V @ 100 Hz 5.643 Vrms 5.757 Vrms 5.9e-3 9.7:1
Square wave 10V @ 100 Hz 9.9 Vrms 10.1 Vrms 8e-3 12.5:1
Square wave 8.0 V @ 100 Hz 7.92 Vrms 8.08 Vrms 7e-3 11.4:1
Agilent E1441A DC Function Offset
Function Output Minimum Measured Maximum M.U. TAR
DC Volts 10.0 Vdc 9.8 Vdc 10.2 Vdc 0.25e-3 >10:1
DC Volts -10.0 Vdc -10.2 Vdc -9.8 Vdc 0.25e-3 >10:1