Agilent Technologies E1441A Oven User Manual


 
Service Procedures 187
Appendix D
3. Send the CAL:VALue <measured Vac> to the function generator.
4. Send the CAL? command to the function generator to initiate the
calibration.
5. Configure SETUP 65 in Table D-22 and repeat steps 3 and 4 for
SETUP 65 through 82 in table D-22.
6. Perform the “Amplitude Flatness Verification” on page 171.
Table D-22. AC Flatness Adjustment Setups
Nominal Output
SETUP FREQUENCY AMPLITUDE Adjustment for:
65 100 kHz 3.0 V rms 100 kHz amplitude flatness
66 500 kHz 3.0 V rms 500 kHz amplitude flatness
67 1 MHz 3.0 V rms 1 MHz amplitude flatness
68 3 MHz 3.0 V rms 3 MHz amplitude flatness
69 5 MHz 3.0 V rms 5 MHz amplitude flatness
70 7 MHz 3.0 V rms 7 MHz amplitude flatness
71 9 MHz 3.0 V rms 9 MHz amplitude flatness
72 10 MHz 3.0 V rms 10 MHz amplitude flatness
73 10.5 MHz 3.0 V rms 10.5 MHz amplitude flatness
74 11 MHz 3.0 V rms 11 MHz amplitude flatness
75 11.5 MHz 3.0 V rms 11.5 MHz amplitude flatness
76 12 MHz 3.0 V rms 12 MHz amplitude flatness
77 12.5 MHz 3.0 V rms 12.5 MHz amplitude flatness
78 13 MHz 3.0 V rms 13 MHz amplitude flatness
79 13.5 MHz 3.0 V rms 13.5 MHz amplitude flatness
80 14 MHz 3.0 V rms 14 MHz amplitude flatness
81 14.5 MHz 3.0 V rms 14.5 MHz amplitude flatness
82 15 MHz 3.0 V rms 15 MHz amplitude flatness