Agilent Technologies E1441A Oven User Manual


 
Agilent E1441A SCPI Command Reference 75
Chapter 3
Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
The function generator calibration is secured when shipped from the factory
and the security code is set to "HP_E1441".
*RST does not change the state.
Example Set the calibration state to unsecured:
CAL:SEC:STAT OFF,HP_E1441
Unsecure function generator calibration.
:SECure:STATe?
CALibration:SECure:STATe? returns a "1" or "0" to show whether the calibration
security state is enabled (1) or disabled (0). The number is sent to the output buffer.
Example Query the calibration security state:
CAL:SEC:STAT?
Query function generator calibration
security state
enter statement
Enter value into computer.
:SETup
CALibration:SETup <0 | 1 | 2 | 3 | . . . | 84> configure the function generator's
internal state for each of the calibration steps to be performed.
:SETup?
CALibration:SETup? queries the calibration setup number. Returns a value
between 0 and 84.
:STRing
CALibration:STRing <quoted string> allows you to record calibration
information about your function generator while CAL:SEC:STAT is OFF. For
example, you can store information such as the last calibration date and/or the next
calibration due date. The calibration message may contain up to 40 characters. The
string is stored in non-volatile memory.
Parameters
Comments The calibration message can contain up to 40 characters.
Calibration security state must be OFF to store a string.
Parameter Name Parameter Type Range of Values Default Units
quoted string discrete alphanumeric none