Agilent Technologies E1441A Oven User Manual


 
Agilent E1441A Application Information 49
Chapter 2
Phase-Lock
Capabilities
(Opt 001)
This section gives an overview of the basic techniques used to program the
Phase-Lock option (option 001). This section is only an overview and does
not give all of the details you will need to write your own application
programs. Refer to “Application Program Examples,” later in this chapter
for more details and examples. Also refer to the programming reference
manual that came with your computer for details on outputting command
strings and entering data.
To Phase Lock to an
External Clock Signal
(requires Option 001)
The front-panel 10 MHz Ref In terminal allows you to synchronize one or
more function generators with an external 10 MHz signal. The following
statements show how to configure a single instrument for synchronization
with an external signal:
APPL:SIN 10E+6, 5.0 Select sine function at 10 MHz
PHAS:ADJ -90 DEG Set phase offset to -90 degrees
PHAS:REF
Set phase reference to zero
Figure 2-11.