Agilent Technologies E1441A Oven User Manual


 
197
Appendix D
Optional Performance Verification Tests
Table D-26. Performance Test Record for the Agilent E1441A (Page 7 of 7)
Agilent E1441A Square Wave Duty Cycle
Function Output Minimum Measured Maximum M.U. TAR
Square wave 1V @ 300 Hz
50% duty cycle
-0.02Vdc 0.02Vdc 25e-6 >10:1
Square wave 1V @ 300 Hz
25% duty cycle
-0.52Vdc -0.48Vdc 25e-6 >10:1
Square wave 1V @ 300 Hz
75% duty cycle
0.48Vdc 0.52Vdc 25e-6 >10:1
Agilent E1441A Distortion
Function Output Minimum Measured Maximum M.U. TAR
Sine wave 1.1V @ 20 kHz reference reference NA
Pass/Fail
2nd Harmonic 40 kHz >70dB NA NA
Pass/Fail
3rd Harmonic 60 kHz >70dB NA NA
Pass/Fail
4th Harmonic 80 kHz >70dB NA NA
Pass/Fail
5th Harmonic 100 kHz >70dB NA NA
Pass/Fail
Sine wave 1.1V @ 100 kHz reference reference NA
Pass/Fail
2nd Harmonic 200 kHz >60dB NA NA
Pass/Fail
3rd Harmonic 300 kHz >60dB NA NA
Pass/Fail
4th Harmonic 400 kHz >60dB NA NA
Pass/Fail
5th Harmonic 500 kHz >60dB NA NA
Pass/Fail
Sine wave 1.1V @ 1 MHz reference reference NA
Pass/Fail
2nd Harmonic 2 MHz >45dB NA NA
Pass/Fail
3rd Harmonic 3 MHz >45dB NA NA
Pass/Fail
4th Harmonic 4 MHz >45dB NA NA
Pass/Fail
5th Harmonic 5 MHz >45dB NA NA
Pass/Fail
Sine wave 1.1V @ 15 MHz reference reference NA
Pass/Fail
2nd Harmonic 30 MHz >35dB NA NA
Pass/Fail
3rd Harmonic 45 MHz >35dB NA NA
Pass/Fail
4th Harmonic 60 MHz >35dB NA NA
Pass/Fail
5th Harmonic 75 MHz >35dB NA NA
Pass/Fail