Agilent E1441A SCPI Command Reference 113
Chapter 3
50% for square waves.
Possible Conflict with Output Frequency: The duty cycle is auto- matically
adjusted if you select a frequency that is not valid with the present duty cycle.
For example, if you set the duty cycle to 70% and then change the frequency to
8 MHz, the function generator will automatically adjust the duty cycle to 60%
(the upper limit for this frequency). A -221, “Settings conflict” error is
generated and the duty cycle is adjusted.
PULSe:DCYCle?
[SOURce:]PULSe:DCYCle? [MIN|MAX] queries the duty cycle setting. Returns a
value in percent.
SWEep:SPACing
[SOURce:]SWEep:SPACing LINear|LOGarithmic selects linear or logarithmic
spacing for the sweep. The default is Linear. [ Stored in volatile memory ]
SWEep:SPACing?
[SOURce:]SWEep:SPACing? queries the sweep mode. Returns “LIN” or “LOG”.
SWEep:TIME
[SOURce:]SWEep:TIME <time>|MIN|MAX sets the number of seconds required to
sweep from the start frequency to the stop frequency. Select from 1 ms to 500
seconds. The default is 1 second.
MIN = 1 ms. MAX = 500 seconds. The sweep time
setting is stored in volatile memory.
Parameters
Comments The number of frequency points in the sweep depends on the sweep time you
select and is automatically calculated by the function generator.
SWEep:TIME?
[SOURce:]SWEep:TIME? [MIN|MAX] queries the sweep time. Returns a value with
units of seconds.
Parameter Name Parameter Type Range of Values Default Units
time numeric 1 ms to 500 sec seconds