Agilent E1441A SCPI Command Reference 95
Chapter 3
Comments The External modulating source is always enabled.
When both sources are enabled (internal-external), the function generator adds
the internal and external modulating signals (the carrier waveform is actually
modulated with two waveforms).
When only the external source is enabled (AM:SOURce EXT, internal source
disabled), the carrier waveform is modulated only by the external waveform.
You apply the external modulating waveform to the AM Modulation (AM
Mod) terminal. The modulation depth is controlled by the signal level present
(5 volts
peak corresponds to 100% modulation).
*RST Condition: BOTH
AM:SOURce?
[SOURce:]AM:SOURce? queries the modulating source. Returns “BOTH” or “EXT”.
AM:STATe
[SOURce:]AM:STATe OFF|ON disables or enables Amplitude Modulation. To
ensure proper operation, you should enable
AM after you have set up the other
modulation parameters. Only one modulation mode can be enabled at a time. When
you enable
AM, the previous modulation mode is turned off.
*RST Condition: 0 (OFF)
AM:STATe?
[SOURce:]AM:STATe? queries the state of AM. Returns “0” (OFF) or “1” (ON).