Chapter 8
Operating Instructions
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 8-25
Figure 8-27: Destination Screens after Activate Prompt
If the REVIEW ALARMS button is pressed the ventilator immediately
displays the screen shown in Figure 8-27(b) appears. If the CANCEL
button is pressed, the ventilator returns to the settings screen for the
selected breath type as shown in Figure 8-27(c).
5. If the operator elects to review alarm settings, the alarm settings
screen for the destination mode is shown (Figure 8-28). All the alarm
limits are grayed out in this screen, because the selected breath type
(PCV), is not yet active. However the grayed out buttons can be used
to adjust values as seems appropriate for the patient.
Figure 8-28: Alarm Limits Screen (PCV active)
Press the Review Alarms button
to go to this screen.
(b)
Press the Cancel button and
return to this screen.
(c)
Press to activate
new breath type.