REF 1057983 A Respironics V200 Ventilator Operator’s Manual 13-89
Chapter 13
2. Attach one end of the RS-232 serial communications cable to the RS-
232 port 1 or 2 on the back of the respiratory profile monitor.
See Figure 13-44.
3. Attach the other end of the RS-232 serial communications cable to
the serial communications port on the back of the ventilator. See
Figure 13-44.
Figure 13-44: RS-232 Communication Ports for the Monitor and the Ventilator
4. Connect the patient circuit to the ventilator.
5. Power up the ventilator if not already operating.
6. Power up and connect the monitor according to the user’s manual for
your Respironics respiratory profile monitor.
Inspect the monitor before powering up.
Press the Operate/Standby key to turn the monitor on.
WARNING: Always turn the ventilator power OFF before connecting additional
equipment.
NOTE: When powering on the ventilator for the first time after installing the
respiratory profile monitor software, the Trending Memory Card must be
reformatted. Press YES when asked the question: