REF 1057983 A Respironics V200 Ventilator Operator’s Manual 6-1
Chapter 6. Connecting Additional Equipment
Communication
Interface
The ventilator provides three communications interfaces: one serial RS-232
port, an analog output port, and a parallel port. (The parallel port is reserved
for use by the Communications Option.)
Figure 6-1: Ports and Outlets on the Rear Panel
WARNING: Always turn the ventilator power OFF before connecting additional
equipment.
CAUTION: To prevent the risk of excessive leakage due to external equipment being
connected to the ventilator via the communication ports, a means for
external separation of the conductive earth paths must be provided.
CAUTION: All equipment used and connected to the ventilator communications
ports (analog, parallel, and serial) must comply with the medical
electrical equipment (IEC601-1) or other applicable standards.
Circuit Breaker
(cover not shown)
Humidifier AC
Outlet
AC Power Cord
AnalogParallel
Remote Alarm
Serial