Chapter 4
Getting Started
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 4-21
Oxygen Source Connection Instructions
Refer to Figure 4-21.
1. Locate the O
2
filter with water trap (1) located in the lower left corner
of the rear panel.
2. Ensure that the O
2
filter and water trap (1) are properly attached to
the ventilator. Inspect the bowl (1) for cracks or potential leaks.
Ensure all connections are tight.
3. Inspect the O
2
hose (2), hose connector (3) and the hose connector
mate (4).
4. Check the oxygen gas supply. Clean, adjust pressure, and drain
condensate from any water traps, filters, or regulators in the O
2
supply
lines.
5. Connect O
2
gas connector (3) to the ventilator O
2
filter with DISS hose
connector (4).
6. If gas source is not already turned on, turn it on and verify that no O
2
gas is leaking.