REF 1057983 A Respironics V200 Ventilator Operator’s Manual 13-111
Chapter 13
Displayed Data While in Speaking Mode, the ventilator switches all volumetric readings from
exhaled to delivered readings. This means that all data displayed based on
exhaled volumes will now be based on delivered volumes. Some data is not
available in Speaking Mode, including data on Patient Leak (circled below in
Figure 9).
Figure 13-65: Active Mode: Displayed Data in the VCV Settings Screen
The following data in the Patient Data screen displays in delivered volumes
(not exhaled volumes):
Tidal Volume
Spontaneous Minute Volume
Total Minute Volume
The following data is not
available in Speaking Mode. Dashed lines appear in
the data box instead of a numeric value:
Patient Leak
F/Vt (Ratio of respiratory rate to tidal volume)
Patient Leak data
is not available in
Speaking Mode