Chapter 10
Care and Maintenance
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 10-9
Removing and Replacing O
2
Input Filter Element
Figure 10-5: Removing and Replacing O
2
Input Filter Element
Refer to Figure 10-5.
1. Disconnect O
2
Hose (1).
2. Unscrew filter housing (3).
3. Unscrew filter element retaining screw (4).
4. Remove and replace filter element (5).
5. Reconnect filter element retaining screw (4).
6. Reinstall filter housing (3).
7. Reconnect O
2
hose (1).
8. Check system for leaks prior to patient use.
1
5
2
3
4