Appendix A
RS-232 Communications Protocol
REF 1057983 A Respironics V200 Ventilator Operator’s Manual Appendix A-3
VCV E-Cycle Type %Flow N/A %FLOW
AUTO◆◆
N/A
VCV %Flow Expiratory
Cycle Setting
25◆◆◆◆ 110 - 80%
VCV Rise Time 0.1◆◆◆ 0.1 0.1 - 0.9 Sec
VCV Oxygen
Concentration Setting
21◆◆◆◆ 1 21 - 100 %
VCV Plateau Setting 0.0◆◆◆ 0.1 0.0 - 2.0 Sec
VCV Apnea Rate
Setting
12◆◆◆◆ 11 - 80BPM
VCV High Inspiratory
Pressure Alarm Limit
Setting
35◆◆◆◆ 1 10 - 105 cmH
2
O
(hPa)
VCV Low Inspiratory
Pressure Alarm Limit
Setting
3◆◆◆◆◆ 13 - 105cmH
2
O
(hPa)
VCV Low PEEP Alarm
Limit Setting
0◆◆◆◆◆ 10 - 35cmH
2
O
(hPa)
VCV Low Mandatory
Tidal Volume Alarm
Limit Setting
0◆◆◆◆◆ 1 0 - 2500 mL
VCV Low Spontaneous
Tidal Volume Alarm
Limit Setting
0◆◆◆◆◆ 1 0 - 2500 mL
VCV High Respiratory
Rate Alarm Limit
Setting
150◆◆◆ 10 - 150BPM
VCV Low Minute
Volume Alarm Limit
Setting
1.00◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0 L
VCV Apnea Interval
Alarm Limit Setting
20◆◆◆◆ 110 - 60sec
PCV Mode Setting A/C◆◆◆ N/A A/C◆◆◆
SIMV◆◆
CPAP◆◆
N/A
PCV Patient Type ADULT N/A ADULT
PED◆◆◆
NEO◆◆◆
PCV Respiratory Rate
Setting
12◆◆◆◆ 11 - 150BPM
PCV Pressure Setting 20◆◆◆◆ 15 - 100cmH
2
O
(hPa)
Ventilation Report (Continued)
Description Example Resolution Range Units Comments
Table A-1: Ventilation Report (Sheet 2 of 10)