Philips V200 Ventilation Hood User Manual


 
Options and Accessories
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 13-41
Chapter 13
Not used  N/A N/A N/A Outputs example
value.
Date FEB231997 N/A N/A N/A 12 character field,
MMMDDYYYY
Mode setting CMV N/A CMV
SIMV
CPAP
N/A Assist/Control =
CMV, Spont/Timed
= SIMV, Spont =
CPAP
Active Respiratory
Rate setting
12.0 0.1 or 1 1.0 – 9.0
10 – 80
BPM Outputs 
when mode is
CPAP or Spont.
VCV Tidal Volume
setting
0.50 0.01 0.05 – 2.50 L
VCV Peak Flow
setting
5 1 3 140 LPM
Oxygen Concentration
setting
100 121 100%
Pressure Trigger
setting
0.5 0.1 0.1 20.0 cm
H
2
O
Outputs value from
active type (VCV or
PCV). When active
type is NPPV then
outputs the VCV
value.
PEEP or EPAP setting 0.0 0.1 0.0 35.0 cm
H
2
O
Uses PEEP setting
from active
ventilation type
(VCV or PCV).
When active type
is NPPV uses
EPAP setting.
VCV Insp. Hold
Setting
0.0 0.1 0.0 2.0 Sec
Not used  N/A N/A N/A Outputs example
value.
Not used  N/A N/A N/A Outputs example
value.
Not used  N/A N/A N/A Outputs example
value.
Not used  N/A N/A N/A Outputs example
value.
VCV Apnea Interval
setting
10 110 60sec
VCV Apnea Tidal
Volume setting (same
as Tidal Volume
setting)
0.50 0.01 0.05 – 2.50 L Outputs VCV Tidal
Volume setting
Ventilation Report
Description Example Resolution Range Units Comments
Table 13-4: Ventilation Report (Sheet 2 of 10)