Warnings, Cautions, and Notes
Disposable or single-patient filters must be discarded between
patients. Do not chemically disinfect or expose single patient use
bacteria filters to ETO gas.
The patient must be disconnected from the ventilator before
entering the Diagnostic Mode since normal ventilation is
suspended.
Do not use a ventilator that has failed SST without verifying
operational readiness by other means. Doing so may place a
patient at risk.
Never initiate SST while the patient is connected to the ventilator.
The high airway pressures generated during SST can injure a
patient.
Never initiate EST while the patient is connected to the ventilator.
The high airway pressures and gas flows generated during EST can
injure a patient.
Do not use a ventilator that has failed EST without verifying
operational readiness by other means. Doing so may place a
patient at risk.
Remove the ventilator from service and contact trained service
personnel if any diagnostic codes appear with the exception of:
1, 3, 2000, 3000, 5000, 5002, 8003, or 8004.
Use of a ventilator that has not passed SST or EST is against the
strongest recommendation of Respironics.
Please contact Respironics Customer Service at 1-800-345-6443
or consult your service manual if any diagnostic codes are
encountered.
When connecting a humidifier to the humidifier outlet (available
only on 100-120 VAC ventilators) allowable leakage current values
may be exceeded.
The use of accessories, cables, and transducers other than those
specified may result in increased EM emissions or decreased
immunity of the system.
We recommend that you use an oxygen monitor that complies with
ISO-7767; Oxygen Monitors or Monitoring Patient Breathing
Mixtures - Safety Requirements. This requirement ensures that
the desired fraction of inspired oxygen (FiO2) is delivered to the
patient.
The batteries (backup battery) in the battery compartment are
non-spillable sealed lead acid. Recycle or dispose of batteries
properly.
Do not connect the DC power cord from the backup battery while
the Respironics V200 is functioning as a ventilator. Always turn
the Power On/Off switch to off ( ).