Chapter 4
Getting Started
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 4-17
Inspiratory Bacteria Filter Installation Instructions
Refer to Figure 4-19.
1. Locate the gas outlet port (1) on the front panel.
2. If the optional O
2
sensor (2) will be used, connect it to the gas outlet
port (1).
3. Remove the inspiratory bacteria filter from the filter package and
inspect for cracks or potential leaks. Discard the filter if it is cracked,
has moisture inside, or is otherwise unserviceable.
4. Some bacteria filters provide an arrow or other mark to indicate the
direction of flow. The flow indicator should be pointed away from the
ventilator, toward the patient circuit connection
Insert inspiratory filter inlet (4) into either
• the optional O
2
sensor (2) using the 22mm connector (3)
or
• the gas outlet port (1) if the optional O
2
sensor (2) is not used.
5. Connect inspiratory limb of patient circuit (5) to bacteria filter (4).
WARNING: To reduce the chance of contamination or infection, always use an
inspiratory and expiratory filter when the ventilator is in operation. Refer to
manufacturer’s instructions and follow institutional infection control
guidelines when replacing the inspiratory and expiratory filter.
WARNING: Do not use anti-static or conductive hoses or conductive patient tubing.
NOTE: Follow institutional infection control guidelines when replacing the
inspiratory or expiratory bacteria filter.
NOTE: When adding attachments or other components or subassemblies to the
breathing system, for example, an HME or humidifier, ensure that the
inspiratory and expiratory resistances (measured at the patient
connection port) do not exceed 6 cmH
2
O (hPa) at a flow of 60 L/min for
adults, 30L/min for pediatrics.