REF 1057983 A Respironics V200 Ventilator Operator’s Manual 13-55
Chapter 13
5. At the end of the maneuver (whether successful or cancelled), the
Stop button changes to Continue. While the Continue button is
displayed, the graph is frozen and can be rescaled. Touch Continue or
exit the screen to unfreeze the display.
6. Touch Continue: the button changes to Start and you can repeat the
maneuver.
Figure 13-30: MIP/P0.1 Screen
WARNING: Perform the MIP maneuver according to instructions. Because machine
assisted breath delivery, apnea detection, and detection of certain patient
alarms are temporarily disabled during a MIP maneuver, close clinical
supervision is recommended.
WARNING: Because PEEP is temporarily set to 0 cmH
2
O during a MIP maneuver, close
clinical supervision is recommended.
MIP
window
P0.1
window
Rescale
button
Start MIP
Press &
Hold
button
Scale
button
Start
P0.1
button