REF 1057983 A Respironics V200 Ventilator Operator’s Manual 13-35
Chapter 13
XHIGH
LEAK
RATE
High Leak
Rate
High leak alarm limit violation. Yellow Alarm.
XLOW EXH
MAND TV
Low Vt
Mandatory
Low mandatory tidal volume limit violation. Yellow
Alarm.
X HIGH EXH
MV
High Minute
Volume
High exhaled minute volume limit violation. Yellow
Alarm.
XLOW EXH
TV
Low Tidal
Volume
Low exhaled tidal volume limit violation. Yellow Alarm.
XLOW EXH
TV-SPNT
Low Vt
Spontaneous
Low spontaneous tidal volume limit violation. Yellow
Alarm.
XINS TI
TOO
LONG
I-Time Too
Long
Maximum inspiratory time exceeded. Yellow Alarm.
XHIGH
RESP
RATE
High
Respiratory
Rate
High respiratory rate limit violation. Yellow Alarm.
XHIGH
O2%
High O2 High measured oxygen concentration. Yellow Alarm.
XLOW
PEEP/
EPAP
Low PEEP Low PEEP limit violation. Yellow Alarm.
XLOW
PEEP/
EPAP
Low EPAP Low EPAP limit violation. Yellow Alarm.
X BACKUP
BATT ON
Battery In Use
indicator on
ventilator
front panel
Ventilator has switched from AC power to Backup
Battery operation. Yellow Alarm.
Inops
X LOSS OF
GAS-SVO
Gas Supplies
Lost-SVO
Oxygen and air sources lost, safety valve open.
XLOW
BACKUP
BATT
Low Backup
Battery
Low Backup Battery voltage.
XALARM
FAILURE
Audible Alarm
Failed
Audible alarm failure.
XAIR
FAULT
Air Source
Fault
Internal air source failure.
XO2 VALVE
STUCK
O2 Valve
Stuck Closed
Defective oxygen valve.
XEXH
VALVE
STUCK
Exp Valve
Stuck Open
Defective exhalation valve.
Available
on
CareNet
CMS
message
V200
equivalent