Philips V200 Ventilation Hood User Manual


 
Chapter 12
12-6 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Technical Specifications
Front Panel Keys
I:E 9.9:1 to 1:99 none 0.1 for 9.9:1-1:9.9;
1 for 1:10-1:99
±10%
Peak Inspiration
Pressure
-20.0 to 130 cmH
2
O
(hPa)
1 for 100-130;
0.1 for -20.0 to 99.9
±10%
End Inspiration
Pressure (Pi End)
-20.0 to 130 cmH
2
O
(hPa)
1 for 100-130;
0.1 for -20.0 to 99.9
±10%
Mean Airway
Pressure
-20.0 to 120 cmH
2
O
(hPa)
1 for 100-120;
0.1 for -20.0 to 99.9
±10%
Delivered O
2
% 0.0 to 110 -- 1 for 100-110;
0.1 for -20.0 to 99.9
±3 vol%
Total Respiratory Rate 0.0 to 150 Bpm 1 for 10-150;
0.1 for 0.0 to 9.9
±10%
Spontaneous
Respiratory Rate
0.0 to 150 Bpm 1 for 10-150;
0.1 for 0.0 to 9.9
±10%
End Exhalation
Pressure (Pe End)
-20.0 to 99.9 cmH
2
O
(hPa)
0.1 ±10%
Patient Leak (Pt Leak) 0.0 to 140 LPM 0.1 for 0.0-99.9
1 for 100-140
Not
specified
a
Percent Patient Trigger
(%Pt Trigger)
0.0 to 99.9 % 0.1 for 0.0-99.9
1 for 100
±10%
a. Due to the variable nature of patient leaks, this parameter is an estimate only.
Patient Data Range, Resolution, Units & Accuracy (Continued)
Patient Data Display Range Units Resolution Accuracy
Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 2 of 2)
Front Panel Keys
Alarm Silence Screen Lock
Alarm Reset Expiratory Hold
100% O
2
Accept
Manual Breath
Table 12-11: Front Panel Keys