Chapter 13
13-100 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Options and Accessories
Initial Neonatal Settings
When the ventilator is first powered on and the neonatal patient type is
selected, the initial startup settings found in Table 13-21: are applied. Each
subsequent startup will display the settings from the last use of the ventilator
of that patient type with the exception of O
2
. The last O
2
setting used on the
ventilator will remain regardless of patient type.
Initial Patient Settings, Ranges
Ventilation Settings Value Ranges
Ventilation Mode PCV - SIMV AC, SIMV, CPAP
Rate 20 BPM 1 - 150 BPM
Pressure 10 cmH
2
O 5 - 100 cmH
2
O
I-Time 0.3 Sec 0.1 - 2.0 Sec
PEEP 3 cmH
2
O0 - 35 cmH
2
O
PSV 0 cmH
2
O 0 - 100 cmH
2
O
I-Trigger 2.0 L/Min 0.3 - 10.0 L/Min
E-Cycle 25% 10 - 80%
Rise-TIme 0.2 Sec 0.1 - 0.5 Sec
O
2
% same as last O
2
setting 21 - 100%
Apnea Rate 20 BPM 1 - 80 BPM
Alarm Settings Value Ranges
High Press 30 cmH2O 10 - 105 cmH2O
Low Insp Press 5 cmH2O 3 - 105 cmH2O
Low Peep 0 cmH2O 0 - 35 cmH2O
Low Vt Mand 10 mL 0 - 300 mL
Low Vt Spont 3 mL 0 - 300 mL
High Rate 150 BPM 0 - 150 BPM
High VE 1.2 L 0.00 - 5.0 L
Low VE 0.3 L 0.00 - 5.0 L
Apnea 20 Sec 10 - 60 Sec
I-Time Too Long sounds after 1.5 Sec of spont
breath
Table 13-21: Initial Neonatal Patient Settings