Philips V200 Ventilation Hood User Manual


 
Appendix A
RS-232 Communications Protocol
REF 1057983 A Respironics V200 Ventilator Operator’s Manual Appendix A-15
PCV I-Trigger Type PRESS N/A PRESS
FLOW◆◆
AUTO◆◆
N/A
PCV Pressure I-Trigger
Setting
2.0◆◆◆ 0.1 0.1 - 20.0 cmH
2
O
(hPa)
PCV Flow I-Trigger
Setting
3.0◆◆◆ 0.1 0.3 - 20.0 LPM
PCV E-Cycle Type %FLOW N/A %FLOW
AUTO◆◆
N/A
PCV %Flow Expiratory
Cycle Setting
25◆◆◆◆ 1 10 - 80 %
PCV Rise Time 0.1◆◆◆ 0.1 0.1 - 0.9 sec
PCV Oxygen
Concentration Setting
21◆◆◆◆ 1 21 - 100 %
PCV Apnea Rate
Setting
12◆◆◆◆ 11 - 80BPM
PCV Apnea Interval
Alarm Limit Setting
20◆◆◆◆ 1 10 - 60 sec
PCV High Inspiratory
Pressure Alarm Limit
Setting
35◆◆◆◆ 1 10 - 105 cmH
2
O
(hPa)
PCV Low Inspiratory
Pressure Alarm Limit
Setting
3◆◆◆◆◆ 13 - 105cmH
2
O
(hPa)
PCV Low PEEP Alarm
Limit Setting
0◆◆◆◆◆ 10 - 35cmH
2
O
(hPa)
PCV Low Mandatory
Tidal Volume Alarm
Limit Setting
0◆◆◆◆◆ 1 0 - 2500 mL
PCV Low Spontaneous
Tidal Volume Alarm
Limit Setting
0◆◆◆◆◆ 1 0 - 2500 mL
PCV High Respiratory
Rate Alarm Limit
Setting
150◆◆◆ 1 0 - 150 BPM
PCV Low Minute
Volume Alarm Limit
Setting
1.00◆◆ 0.01 for
0.00 to
9.99
0.1 for
10.0 to
60.0
0.00 - 60.0 L
Pressure Control Ventilation Settings Report (Continued)
Description Example Resolution Range Units Comments
Table A-3: Pressure Control Ventilation Settings Report (Sheet 2 of 3)