Chapter 11
11-2 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Diagnostics
Entering Diagnostic
Mode
To enter Diagnostic Mode, hold down the ALARM RESET and 100% O
2
keys
for approximately 5 seconds while you turn on the ventilator. A message
appears on the ventilator prompting you to confirm that the patient is
disconnected before entering Diagnostic Mode (Figure 11-1). Press OK to
enter Diagnostic Mode.
Figure 11-1: Entering Diagnostic Mode
Once you’ve entered Diagnostic Mode, you can select any of diagnostic
functions by pressing its button. Change values in Diagnostic Mode the same
way you change ventilator settings.