Chapter 8
Operating Instructions
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 8-29
Monitor Screen The monitor screen (Figure 8-34) is the default screen, and is automatically
displayed if the screen has not been touched for 15 minutes. The screen can
also be viewed by pressing the MONITOR button. The screen is shown in
Figure 8-34. The patient data and settings displayed on this screen depend on
the active breath type (Refer to Table 8-8: “Monitor Screen Settings
Displayed” and Figure 8-34.)
Spont VE Spontaneous Minute Volume
The spontaneous ventilation
normalized to one minute.
Compliance compensated (if
enabled).
0.00 to 99.9 L
(BTPS)
0.01 for 0.00 to
9.99
0.1 for 10.0 to 99.9
Total VE Exhaled Minute Volume
The total volume exhaled by the
patient in one minute.
Compliance compensated (if
enabled.)
0.00 to 99.9 L
(BTPS)
0.01 for 0.00 to
9.99
0.1 for 10.0 to 99.9
% O
2
Delivered O
2
One second average of O
2
sensor reading (if O
2
sensor is
installed).
0.0 to 110% 0.01 for 0.00 to
9.99
1 for 100 to 110
Spont Rate Spontaneous Respiratory Rate
The respiratory rate of
spontaneous breaths.
0.0 to 150
Bpm
0.1 for 0.0 to 9.9
1 for 10 to 150
Total Rate Total Respiratory Rate
The respiratory rate for all
breaths.
0.0 to 150
Bpm
0.1 for 0.0 to 9.9
1 for 10 to 150
F/V
t
Rapid Shallow Breathing Index
The ratio of the respiratory rate
to exhaled tidal volume for
spontaneous breaths.
0 to 150 Bpm/
L
1
I:E Ratio I:E Ratio
The ratio of inspiratory time to
expiratory time.
9.9:1 to 1:99 0.1 for 9.9:1 to
1:9.9
one (1) for 1:10 to
1:99
NOTE: Pt. Leak only appears on Patient Data block on Settings screen.
Patient Data Definitions: Range, Units & Resolution (Continued)
Parameter Description Display Range Resolution
Table 8-7: Patient Data Definitions: Range, Units & Resolution (Sheet 2 of 2)