Philips V200 Ventilation Hood User Manual


 
Chapter 13
13-14 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Options and Accessories
Assembly Instructions 1. Lock the front wheels of the ventilator.
Figure 13-9: Locking Ventilator Wheel
WARNING: The Oxygen Manifold is for oxygen use only.
WARNING: Maximum inlet pressure: 90 PSIG
Minimum inlet pressure: 50 PSIG at 200 SLPM
WARNING: Keep all hoses within the limits of the ventilator footprint to prevent
a tripping or snag hazard.
CAUTION: Use this kit with the Respironics V200 Ventilator only.
CAUTION: Do not use the oxygen manifold or oxygen hoses to move the ventilator.