Philips V200 Ventilation Hood User Manual


 
Chapter 12
Technical Specifications
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 12-9
Connectors
Alarm
Settings
High Inspiratory
Pressure
10 to 105 cmH
2
O (hPa)
Low Inspiratory
Pressure
3 to 105 cmH
2
O (hPa)
Low PEEP
Pressure
0 to 35 cmH
2
O (hPa)
High Respiratory
Rate
0 to 150 Bpm
Low Exhaled
Mandatory Tidal
Volume
0 mL to 2500 mL
Low Exhaled
Spontaneous Tidal
Volume
0 mL to 2500 mL
Low Exhaled
Minute Volume
0 to 60 Lpm
High Exhaled
Minute Volume
0 to 60 Lpm
High Leak (NPPV
mode only)
0 to 60 Lpm
Alarm
Status
Indicators
Normal Indicator
High Urgency Alarm (flashing red)
Medium/Low Urgency Alarm (flashing yellow/steady yellow)
Alarm Silence
Safety Valve Open
Ventilator Inoperative (steady red)
Screen Locked
Alarms (Continued)
Table 12-17: Alarms (Sheet 2 of 2)
Connectors
Inspiratory Limb Connector Gas outlet port: 22 mm conical male
Expiratory Limb Connector Gas return port (on expiratory filter):
Proprietary barb fitting
NOTE: The gas return port on the ventilator is a
cylindrical port which requires mating to a
specified expiratory filter to seal the expiratory
limb.
Oxygen Inlet DISS male, DISS female, NIST, Air Liquide, or SIS Fitting
(depending on country and configuration).
Table 12-18: Connectors