Appendix A
Appendix A-4 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
RS-232 Communications Protocol
PCV Inspiratory Time
Setting
1.00◆◆ 0.01 0.1 - 9.9 sec
PCV PEEP Setting 0◆◆◆◆◆ 10 - 35cmH
2
O
(hPa)
PCV Pressure Support
Setting
0◆◆◆◆◆ 10 - 100cmH
2
O
(hPa)
PCV I-Trigger Type PRESS◆ N/A PRESS◆
FLOW◆◆
AUTO◆◆
N/A
PCV Pressure I-Trigger
Setting
2.0◆◆◆ 0.1 0.1 - 20.0 cmH
2
O
(hPa)
PCV Flow I-Trigger
Setting
3.0◆◆◆ 0.1 0.3 - 20.0 LPM
PCV E-Cycle Type %FLOW◆ N/A %FLOW◆
AUTO◆◆
N/A
PCV %Flow Expiratory
Cycle Setting
25◆◆◆◆ 110 - 80%
PCV Rise Time 0.1◆◆◆ 0.1 0.1 - 0.9 sec
PCV Oxygen
Concentration Setting
21◆◆◆◆ 1 21 - 100 %
PCV Apnea Rate
Setting
12◆◆◆◆ 11 - 80BPM
PCV Apnea Interval
Alarm Limit Setting
20◆◆◆◆ 110 - 60sec
PCV High Inspiratory
Pressure Alarm Limit
Setting
35◆◆◆◆ 1 10 - 105 cmH
2
O
(hPa)
PCV Low Inspiratory
Pressure Alarm Limit
Setting
3◆◆◆◆◆ 13 - 105cmH
2
O
(hPa)
PCV Low PEEP Alarm
Limit Setting
0◆◆◆◆◆ 10 - 35cmH
2
O
(hPa)
PCV Low Mandatory
Tidal Volume Alarm
Limit Setting
0◆◆◆◆◆ 1 0 - 2500 mL
PCV Low Spontaneous
Tidal Volume Alarm
Limit Setting
0◆◆◆◆◆ 1 0 - 2500 mL
PCV High Respiratory
Rate Alarm Limit
Setting
150◆◆◆ 10 - 150BPM
Ventilation Report (Continued)
Description Example Resolution Range Units Comments
Table A-1: Ventilation Report (Sheet 3 of 10)