Philips V200 Ventilation Hood User Manual


 
Chapter 4
4-20 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Getting Started
Oxygen Source
Connection
Figure 4-21: O
2
Gas Connections and Filter
WARNING: All oxygen connections should be carefully inspected to ensure that leaks
are not present. Excessive leaks can result in higher than normal ambient
oxygen concentrations and create a potentially hazardous oxygen-enriched
environment.
WARNING: Worn/frayed oxygen hoses or oxygen hoses contaminated by hydrocarbon
greases or oils should not be used since an oxygen leak or intense fire
could result.
WARNING: Care in the routing of the oxygen inlet hose should be exercised to ensure it
is not exposed to mechanisms that could cause damage by cutting or
heating/melting.
CAUTION: The ventilator oxygen filter should be replaced annually as a part of
preventive maintenance.
NOTE: The ventilator should only be connected to an appropriate medical
grade 100% O
2
gas source capable of delivering a regulated 40 to 90
PSIG (276-620 kPa).
NOTE: The ventilator is shipped with the appropriate gas fittings and hoses for
the intended environment, i.e. DISS (U.S.A. and Canada), Ohmeda
(Germany), NIST (UK), Air Liquide (France), SIS (Australia).