12-10 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Technical Specifications
Filters
Measuring and
Display Devices
AC Power and Battery
Indicators
Filters
Inspiratory See filter instruction sheets for complete specification.
Bacterial Filter Efficiency: 99.999+%
Viral Filtration Efficiency: 99.99+%
Expiratory See filter instruction sheets for complete specification.
Typical efficiency: 99.97% for nominal particle size of 0.3 micro meter (micron)
at 100L/min flow.
Table 12-19: Filters
Measuring and Display Devices
Pressure Measurements Type Solid State
Sensing Position Inhalation output port
Exhalation input port
Measurements Patient airway pressure
Volume Measurements Type Hot Film Anemometer flow integrated
Sensing Position O
2
, air and exhalation
Measurements O
2
flow, air flow and exhalation flow
Oxygen Measurements Type Zirconia solid electrolyte
Sensing Position Internal enclosure
Measurements Oxygen concentration (range 0 to 95%)
Type Galvanic cell
Sensing Position Inspiratory limb of VBS
Measurements Delivered O
2
% (range 0 to 110%)
Settings, Alarms and
Data Display
Type TFT liquid crystal touch screen
Table 12-20: Measuring and Display Devices
AC Power & Battery Indicators
Loss of Power (audible only) Backup Battery Low
External Battery In Use Backup Battery In Use
Backup Battery Charging Mains
Table 12-21: AC Power & Battery Indicators