Philips V200 Ventilation Hood User Manual


 
Appendix A
RS-232 Communications Protocol
REF 1057983 A Respironics V200 Ventilator Operator’s Manual Appendix A-5
PCV Low Minute
Volume Alarm Limit
Setting
1.00◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0 L
PCV High Minute
Volume Alarm Limit
Setting
60.0◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0 L
NPPV Mode Setting SPONT/T N/A SPONT/T
SPONT◆◆
NPPV Patient Type ADULT N/A ADULT
PED◆◆◆
NEO◆◆◆
NPPV Respiratory
Rate Setting
12◆◆◆◆ 11 - 80BPM
NPPV EPAP Setting 5◆◆◆◆◆ 12 - 25cmH
2
O
(hPa)
NPPV IPAP Setting 5◆◆◆◆◆ 12 - 35cmH
2
O
(hPa)
NPPV Inspiratory Time
Setting
1.0◆◆◆ 0.1 0.1 - 9.9 sec
NPPV Rise Time 0.1◆◆◆ 0.1 0.1 - 0.9 sec
NPPV I-Trigger Type AUTO N/A FLOW◆◆
AUTO◆◆
N/A
NPPV Flow I-Trigger
Setting
3.0◆◆◆ 0.1 0.5 - 20.0 LPM
NPPV E-Cycle Type AUTO◆◆ N/A FLOW◆◆
AUTO◆◆
N/A
NPPV %Flow
Expiratory Cycle
Setting
25◆◆◆◆ 110 - 80%
NPPV Oxygen
Concentration Setting
21◆◆◆◆ 1 21 - 100 %
NPPV Apnea Rate
Setting
12◆◆◆◆ 11 - 80BPM
NPPV Low Inspiratory
Pressure Alarm Limit
Setting
3◆◆◆◆◆ 13 - 105cmH
2
O
(hPa)
NPPV Low EPAP
Alarm Limit Setting
0◆◆◆◆◆ 10 - 25cmH
2
O
(hPa)
NPPV Low Tidal
Volume Alarm Limit
Setting
0◆◆◆◆◆
1
0 - 2500 mL
Ventilation Report (Continued)
Description Example Resolution Range Units Comments
Table A-1: Ventilation Report (Sheet 4 of 10)