Chapter 8
Operating Instructions
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 8-33
7. Push USER CONFIG to check time, time format, date and altitude and
set compliance compensation as required (Figure 8-37).
8. Press START SST to begin Short Self Test, and follow the screen
prompts. Use a cap to plug the patient port of the patient circuit wye
at the prompt.
9. During SST, as each test is performed, the ventilator displays test
results (Figure 8-37). Do not proceed until the ventilator completes
SST without failures.
Figure 8-37: SST Results
NOTE: If time is found to be incorrect more than once in the preoperational
procedure, an internal battery may have to be replaced. Contact
qualified service personnel or call Respironics Customer Service at
1-800-345-6443.
Any failures
detected during
SST are listed
here.
Push USER
CONFIG to set
time, date,
altitude,
compliance
compensation
enable, and time
format.
WARNING: You will be warned if the compliance is 9.0 ml/cmH
2
O (hPa) or larger.
Patients should not be put on a patient circuit that does not meet this
requirement.