Options and Accessories
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 13-43
Chapter 13
Measured peak
inhalation pressure
0.0 0.1 or 1 0.0 – 99.9
100 – 130
cm
H
2
O
Outputs
when ventilation
mode is CPAP or
Spont and No
Apnea alarm is
present.
Measured mean
airway pressure
0.9 0.1 or 1 0.0 – 99.9
100 - 130
cm
H
2
O
Measured end
inhalation pressure
7.7 0.1 or 1 0.0 – 99.9
100 – 130
cm
H
2
O
Outputs
when
Ventilation type is
PCV or NPPV, or
Ventilation mode is
CPAP or Spont, or
VCV Insp. Hold
setting = 0.
Measured I:E ratio 0.2 0.1 0.1 – 9.9
10 - 99
N/A Fractional
representation of
1:X.X with one (1)
decimal place,
e.g., 4:1 = 0.3
High Inhalation
Pressure Alarm
setting
20 110 – 105cm
H
2
O
Outputs high
inhalation pressure
setting for active
ventilation type
(PCV or VCV) or
IPAP + 10 if NPPV
is active.
Low Inhalation
Pressure Alarm
setting
3 13 – 105cm
H
2
O
Outputs Low
Inhalation
Pressure setting
for active
ventilation type.
Low PEEP or Low
EPAP Alarm setting
0 1 0 – 35 cm
H
2
O
Outputs Low PEEP
from active type
(VCV or PCV).
When active type
is NPPV uses Low
EPAP.
Ventilation Report
Description Example Resolution Range Units Comments
Table 13-4: Ventilation Report (Sheet 4 of 10)